AN ANALYSIS OF ORTHODONTIC MINISCREW STABILITY IN EXPERIMANTAL AND CLINICAL CASES

https://doi.org/10.25587/SVFU.2021.23.2.004

Abstract

The aim of this research was an experimental and clinical study measuring stability of the orthodontic miniscrews “Turbo” (Russia) and “BioRay” (Taiwan). Four self-drilling orthodontic miniscrews of each manufacturer, i.e. “BioRay”, Taiwan, (10-mm length, 2-mm diameter) and “Turbo”, Russia, (9-mm length, 2-mm diameter), a total of 8 items, were inserted into native pig mandible sample. Their stability was estimated by torques using a dynamometer (Zahoransky AG, Germany). In the clinical study, 62 orthodontic miniscrews were used: 16 “Turbo” (Russia) and 46 “BioRay” (Taiwan). Their stability was clinically evaluated by the criterion of the presence or absence of mobility of the orthodontic miniscrew after the application of the load. It was established that the orthodontic miniscrew “Turbo” (Russia) were not inferior to “BioRay” (Taiwan) miniscrews in the primary stability, stability after 7 days under loading and in resistance to the load in clinical case. In this regard, Russian orthodontic mini-screws can be successfully applied in clinical practice as well as foreign counterparts for studies of primary stability and loads resistance in clinical conditions.

About the Authors

N. V. Popova
National Medical Research Center “Research Institute of Dentistry and Maxillofacial Surgery” of the Ministry of Health of the Russian Federation
Russian Federation


O. I. Arsenina
National Medical Research Center “Research Institute of Dentistry and Maxillofacial Surgery” of the Ministry of Health of the Russian Federation
Russian Federation


I. Y. Lebedenko
Head of the Laboratory of Materials Science
Russian Federation


F. S. Rusanov
National Medical Research Center “Research Institute of Dentistry and Maxillofacial Surgery” of the Ministry of Health of the Russian Federation
Russian Federation


E. A. Khvorostenko
National Medical Research Center “Research Institute of Dentistry and Maxillofacial Surgery” of the Ministry of Health of the Russian Federation
Russian Federation


N. V. Glukhova
National Medical Research Center “Research Institute of Dentistry and Maxillofacial Surgery” of the Ministry of Health of the Russian Federation
Russian Federation


References

1. Effects of thread depth, taper shape, and taper length on the mechanical properties of mini-implants / Jenny Zwei-Chieng Chang, Yi-Jane Chen, Yuan-Yi Tung, Yu-Ying Chiang, Eddie Hsiang-Hua Lai, Weng-Pin Chen, and Chun-Pin Lin.// Am J Orthod Dentofacial Orthop.-2012.-141-P.279-88. https://doi.org/10.1016/j. ajodo.2011.09.008

2. Influence of geometric design characteristics on primary stability of orthodontic miniscrews/ Eman Saad Radwan , Mona A Montasser , Ahmed Maher. // Am J Orthod Dentofacial Orthop-2018.-79(3)-Р.191-203. https://doi.org/10.1007/s00056-018-0131-7

3. Effects of size and insertion angle of orthodontic mini-implants on skeletal anchorage / Ufuk Tatli, Mohammed Alraawi, and Mustafa Serdar Toroglu. // Am J Orthod Dentofacial Orthop -2019.-156-Р.220-8. https://doi.org/10.1016/j.ajodo.2018.08.026

4. Insertion torque and Periotest values are important factors predicting outcome after orthodontic miniscrew placement / Takashi Watanabe, Ken Miyazawa, Takuya Fujiwara, Misuzu Kawaguchi, Masako Tabuchi, Shigemi Goto.// Am J Orthod Dentofacial Orthop -2017.-152-Р.483-8. https://doi.org/10.1016/j.ajodo.2017.01.026

5. Effect of cortical bone thickness and density on pullout strength of mini-implants: An experimental study/ Fatma Kübra Erbay Elibol , Emre Oflaz , Emine Buğra , Metin Orhan , Teyfik Demir.// Am J Orthod Dentofacial Orthop- 2020.-157(2)-Р.178-185. https://doi.org/10.1016/j.ajodo.2019.02.020

6. Effects of cortical bone thickness and trabecular bone density on primary stability of orthodontic mini-implants/ Chin-YunPan, Pao-HsinLiu, Yu-ChuanTseng, Szu-TingChou, Chao-YiWu, Hong-PoChang. // Journal of Dental Sciences -2019.-14-Р.383-388 https://doi.org/10.1016/j.jds.2019.06.002

7. Insertion angle impact on primary stability of orthodontic mini-implants/ Benedict Wilmes , Yu-Yu Su, Dieter Drescher. // Angle Orthodontist-2008.-78(6)-Р.1065-1070. https://doi.org/10.2319/100707-484.1

8. Pre-disposing factors for orthodontic mini-implant failure defined by bone strains in patient-specific finite element models/ Albogha MN, Kitahara T, Todo M, Hyakutake H, Takahashi I.//Annals of Biomedical Engineering-2016.-44-Р.2948-56. https://doi.org/10.1007/s10439-016-1584-8


Review

For citations:


Popova N.V., Arsenina O.I., Lebedenko I.Y., Rusanov F.S., Khvorostenko E.A., Glukhova N.V. AN ANALYSIS OF ORTHODONTIC MINISCREW STABILITY IN EXPERIMANTAL AND CLINICAL CASES. Vestnik of North-Eastern Federal University. Medical Sciences. 2021;(2):32-38. (In Russ.) https://doi.org/10.25587/SVFU.2021.23.2.004

Views: 82


Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 License.


ISSN 2587-5590 (Online)